ECG
Lead Selection: LI, LII, LIII
Patient Cable: 6-Pin AAMI Standard
Ground Isolation: 4 kV rms, 5.5 kV peak
CMRR: 90 dB
Input Impedance: 20 M at 10 Hz with patient cable
Defibrillator Protection: Protected against 360 J discharge
Leakage Current: <10 µA at 120 V ac, 60 Hz
Electrosurgical Interference Protection: Standard
Gain Selections: 5, 10, 20, 40 mm/mV
Cardiotach
Range: 15 to 300 bpm
Accuracy: ±1%
Resolution: 1 bpm
Sensitivity: 300 µV peak
Pace Pulse Rejection Width: 0.1 to 2 ms at ±2 to ±700 mV
Tall T Wave Rejection: Rejects T waves R wave
Alarms
High Rate: Adjustable in 5 bpm increments
Low Rate: Adjustable in 5 bpm increments
Asystole: R to R interval >6 seconds
Lead Off: Detached Lead or Offset Potential >0.5 V
Display
Type: Active Matrix TFT Color LCD (640x480)
Trace: Single Trace with Freeze Mode
Screen Size: 13.25 cm x 9.94 cm, 16.5 cm (6.5 in.) diagonal
Sweep Speed: Moving Bar Selectable @ 25 or 50 mm/sec
Configurations:
3000 - without recorder
3000R - with recorder
| | Mechanical
Size: Height: 6.70 in. (17.2 cm)
Width: 9.25 in. (23.5 cm)
Depth: 9.21 in. (23.4 cm)
Weight: 6.5 lb. (2.9 kg)
Environmental
Operating Temperature Range: 5°C to 35°C
Storage Temperature Range: -5°C to 55°C
Relative Humidity: 0-90% non-condensing
Power Requirements
Input Range: 85 to 264 V ac, 47 to 63 Hz; +10% -15%
Maximum ac Power Consumption: 35 VA
Filter mode: Auto sense set of filter to 50 or 60 Hz
Regulatory
Unit meets or exceeds the specifications for the AAMI EC/13 Cardiac Monitor
Standard. UL-60601-1, Can/CSA 601.1 No. M90, CDN MDR (CMDCAS),
IEC 60601-1, IEC 60601-2-25, IEC 60601-2-27, ISO 13485:2003, FDA/CGMP, and MDD 93/42/EEC.
CE 0413
Warranty
One Year (parts and labor)
Recorder Option
Number of Channels: 1
Recording Waveform: ECG
Chart Speed: 25 or 50 mm/sec
Channel Width: 40 mm
Mark: Recorder marks the Peak location and the duration of trigger pulse
Chart paper: Standard 50mm ECG thermal paper 2" Dia.Max.
Communications Options
RS 232
(Specifications subject to change without notice)
|