Technical Specifications

 


img Product Brochure

home Overview

 Technical Specs

Available Accessories



Physical characteristics
  • Dimensions: 9” x 7” x 4.5”
  • Weight: 5 lbs.
Environmental
  • Operating Temperature: Range: 5°C to 40°C
  • Storage Temperature Range: -5°C to 55°C
  • Relative Humidity: 0-90% non-condensing
  • Atmospheric Pressure: 500-1060 mbar
  • Protection against ingress of fluids: IPX1 – Ordinary (without protection against ingress of water)
Display
  • 6.5” Color VGA Touch screen LCD (640X480)
  • Viewing angle ≤ 170 degrees
  • Brightness at least 300 nits and backlight cold cathode fluorescent.
Power Supply
  • Input Voltage: 90 – 264V AC
  • Input Frequency: 47 – 63 Hz
  • Input Current Full Load Low Line: 1.5A RMS max
  • Input Leakage Current: < 250µA
  • Output Power: 60W
  • Output Voltage: 12V ± 3%
  • Output Current: 5A max; 0.1A min
  • Output Rise Time: < 100mS
Battery
  • Minimum 3 hours runtime w/on-screen Indicator
  • Removable and rechargeable battery supply
System Features
  • Software easily upgradable
  • Real-Time System Configuration
  • Patient Quick Disconnect (EtCO2 Module)
  • Nurse call Support
  • Night Dimmer
System Audio
  1. High Priority alarm audio
  2. Informational audio
  3. Button key Click
  Wi Fi 802.11 module
  • Communication: TCP and UDP to serial 115.2KB
  • Wireless Interface: 802.11 b/g with WEP, WPA
  • Network Comm: ARP, UDC, TCP, ICMP, SNMP
  • Firmware Updates: TFTP
  • Supply Voltage: 3.3V ±0.15V
  • Operating Temp: 0 to 70 deg C
  • Frequency Range: 2.4 – 2.484 GHz
  • Output Power: 14 ±2 dBm
  • Receiver Sensitivity: > -69 dBm
  • Security: WEP /128,WPA/802.11i
  • Antenna: either internal or external to module
SPO2 Module
  • SpO2 Range: 0 – 100%
  • Accuracy: ± 3% (DS-100A)
  • Resolution: 1%
  • Averaging: 2, 8, or 16 seconds.
  • User selectable
  • Pulse Rate Range: 30 – 300 bpm
  • Pulse Rate Accuracy: ± 1% of full scale
  • Pulse Rate Averaging: 8 seconds
ETCO2 Module
  • ETCO2 Range: 0 – 99mm Hg / 0 – 13% Vol
  • ETCO2 Accuracy: ± 2mmHg from 0 - 38mm Hg
  • ±5% from 39 – 76mm Hg
  • ±8% from 77 – 99mm Hg
  • Breath Rate Range: 0 -150 bpm
  • Breath Rate Accuracy: ± 1%
Regulatory Compliance
Complies with both domestic and international regulatory standards and guidelines relevant to medical devices including:
  • IEC 60601-1: Medical Electrical Equipment, part 1: General Requirement for Safety
  • ISO 14971: Medical Device Risk Analysis
  • IEC 1041: Information Supplied by the Manufacturer with Medical Devices
  • IEC 60601-1-2: Electric Magnetic Compatibility
  • IEC 60601-1-4: Programmable Medical Systems
  • ISO 21647: Performance of Respiratory Gas Monitors
  • ISO 9919: Safety and Performance of Pulse Oximeter for Medical Use
(Specifications subject to change without notice)